1 Scope
This European Standard specifies a continuous measurement method for the determination
of the concentration of carbon monoxide present in ambient air based on the non-dispersive infrared spectroscopic measuring principle. This standard
describes the performance characteristics and sets the relevant minimum criteria required to select an appropriate non-dispersive
infrared spectroscopic analyser by means of type approval tests. It also includes the evaluation of the suitability of an analyser for use in a specific fixed site so as to meet the data quality requirements as specified
in Annex I of Directive 2008/50/EC [1] and requirements during sampling, calibration and quality assurance for use.
The method is applicable to the determination of the mass concentration of carbon
monoxide present in ambient air up to 100 mg/m 3 carbon monoxide. This concentration range represents the certification range for the type approval test.
NOTE 1 Other ranges may be used depending on the levels present in ambient air .
NOTE 2 When the standard is used for other purposes than for measurements required by Directive
2008/50/EC, the ranges and uncertainty requirements may not apply.
The method covers the determination of ambient air concentr...