1 Scope
This document provides specific requirements for the basic safety and essential performance
for oro - tracheal and naso-tracheal tubes and tracheal tube connectors , tracheal tubes with walls reinforced with metal or plastic, tracheal tubes with shoulders , tapered tracheal tubes , tracheal tubes with means for suctioning, monitoring or delivery of drugs or other gases, and the
many other types of tracheal tubes devised for specialized applications.
Tracheobronchial (including endobronchial) tubes (see ISO 16628 ), tracheostomy tubes (see ISO 5366 ), and supralaryngeal airways (see ISO 11712 ) are excluded from the scope of this document.
Tracheal tubes intended for use with flammable anaesthetic gases or agents, lasers, or electrosurgical
equipment are outside the scope of this document.