Overview
EN ISO 2556:2000 defines the manometric method for determining the gas transmission rate (GTR) of plastic films and thin sheets under atmospheric pressure. The procedure measures the volume of gas passing through a specimen per unit area, time and pressure difference (expressed in cm3/m2·d·atm). The standard covers the principle, required apparatus, test specimens, procedure, calculation and reporting.
Key practical points:
- Measurement range: 1 to 20 000 cm3/m2·d·atm.
- Test gases: air or other gases may be used.
- Typical test environment: controlled temperature (e.g., 23 ± 2 °C) and specified conditioning per ISO/R 291.
Key Topics
-
Principle: The specimen separates two chambers. Chamber 1 contains the test gas at atmospheric pressure; chamber 2 is evacuated and sealed. Gas permeating the specimen increases the pressure in chamber 2, which is measured with a manometer (capillary/mercury system) as a function of time.
-
Apparatus components:
- Metal two-part test cell (upper gas chamber and lower test plate forming chamber 2).
- Calibrated capillary tube (manometer) and U-tube connected to a mercury reservoir.
- Vacuum pump capable of reducing chamber 2 to approximately 13 Pa (0.13 mbar).
- Filter paper or sintered disc support for the specimen; rubber gasket to seal the cell.
- Thermometer and mercury barometer for temperature and atmospheric pressure records.
-
Specimens & conditioning:
- Specimens should be representative discs large enough to overlap the gasket (typically ~80 mm diameter) and free from visible defects.
- At least three specimens shall be tested. Hygroscopic materials must be conditioned to the selected standard atmosphere.
-
Calculation: The standard gives the formula for converting the manometer rise (dh/dt) into the gas transmission rate, accounting for capillary geometry, chamber volume, temperature (T), barometric pressure (H) and apparatus correction coefficient.
Applications
- Barrier performance assessment for food, pharmaceutical and medical packaging materials.
- Quality control during production of films and thin sheets to detect changes in permeability.
- Material development and comparison of barrier properties across polymer formulations and laminates.
Benefits include reproducible measurement of permeability under atmospheric pressure and adaptability to different test gases and humidity conditions.
Related Standards
- ISO/R 291 (conditioning atmospheres for plastics) - referenced for specimen conditioning and selected temperatures.
- Associated national and European implementations of ISO 2556 (EN ISO 2556).
For test reports, include sample description and thickness, number of specimens, apparatus details, test gas and humidity, test temperature, individual and mean transmission rates, and any conditioning performed.