Overview
ISO 11083:1994 specifies a spectrometric method for determining chromium(VI) in water using 1,5-diphenylcarbazide. This internationally recognized standard focuses on accurate measurement of dissolved hexavalent chromium concentrations within the range of 0.05 mg/L to 3 mg/L. The method can be extended to higher concentrations through sample dilution. Chromium(VI) is an important contaminant due to its toxicity and environmental impact, making precise water quality testing essential.
The method involves the formation of a red-violet chromium-1,5-diphenylcarbazone complex that is measured spectrometrically at wavelengths between 540 nm and 550 nm. Sample pretreatment steps stabilize chromium valency states and address interferences from other substances to ensure reliable results.
Key Topics
-
Scope and Application
Applicable for dissolved chromium(VI) in surface waters, effluents, and industrial wastewater, with adaptable concentration ranges.
-
Principle
Chromium(VI) reacts with 1,5-diphenylcarbazide to form a colored complex, measured photometrically.
-
Reagents and Solutions
Use analytical-grade reagents such as sodium hydroxide, phosphoric acid solutions, aluminium sulfate, sulfite solution, and 1,5-diphenylcarbazide in acetone.
-
Sample Collection and Pretreatment
Samples are collected in glass bottles, buffered, and adjusted for pH. Pretreatment steps remove or neutralize oxidizing/reducing agents that could affect chromium valency.
-
Interferences
The method addresses interference from metals like lead, barium, silver, molybdenum, mercury, vanadium, and iron by employing selective precipitation and filtration. Special care is taken with samples containing oxidizing or reducing substances to prevent chromium valency changes.
-
Apparatus
Requires a photometer or spectrometer with 10-50 mm path length cells, membrane filtration devices, pH meters, and gas flow controls for air purging.
-
Calibration and Measurement
Calibration curves are created using chromium(VI) standards, with absorbance measured after 5-15 minutes of complex formation. Blank and sample measurements are taken concurrently to correct for turbidity and background absorption.