Overview
ISO 13093:2023 defines a laboratory method for the determination of carbon in titanium and titanium alloys using an infrared absorption method after combustion in a high‑frequency induction furnace under an oxygen atmosphere. The standard covers analysis for carbon contents between 0.003 % and 0.050 % (mass fraction) and specifies required procedures for sample preparation, instrument preparation, blanking, calibration and calculation.
Key topics and technical requirements
- Principle: combust the test portion in an induction furnace in pure oxygen; carbon is converted to CO2 and/or CO and quantified by infrared (IR) absorption.
- Scope of measurement: applicable for carbon from 0.003 % to 0.050 %.
- Reagents & gas: high‑purity oxygen, moisture and CO2 absorbents (e.g., magnesium perchlorate), and accelerators (tungsten, iron or tin) as recommended by the instrument manufacturer.
- Apparatus: high‑frequency induction furnace with IR absorption analyzer, high‑purity ceramic crucibles (pre‑ignited), and gas filters as per manufacturer guidance.
- Sampling & sample prep: use millings/drillings uniform in size (≤5 mm), dry and free from oil/scale; clean with ethanol or acetone.
- Test portion & replication: weigh 200–600 mg (nearest 1 mg); perform determinations at least in duplicate (routine single determinations allowed by agreement).
- Blank tests: perform duplicate blanks using the accelerator only; mean blank must be ≤ 0.001 % and difference between blanks ≤ 5 µg.
- Calibration: use titanium or titanium‑alloy certified reference materials (CRMs/RMs) covering the top, middle and bottom of the method scope; compute an equivalence factor (instrument response → mg C).
- Calculation: subtract blank, apply equivalence factor and sample mass to report carbon as % (mass fraction). Modern analyzers often output %C directly.
- Precision: interlaboratory study (12 labs) yielded concentration‑dependent precision. For example, at 0.003 % C the repeatability r ≈ 0.00074 % and overall reproducibility (smoothed) ≈ 0.0017 %.
Applications and users
This standard is intended for:
- Quality control and materials testing laboratories analyzing titanium and titanium alloys.
- Aerospace, medical‑implant, and specialty metallurgy manufacturers where tight carbon control influences mechanical properties and performance.
- Research labs and certification bodies requiring traceable, validated measurement protocols and CRMs.
ISO 13093:2023 supports consistent, comparable carbon measurements for process control, acceptance testing and regulatory compliance.