1 Scope
This Technical Report gives the detailed evaluation of uncertainty for volume measurements according to the Guide to the Expression of Uncertainty in Measurement (GUM) [1]. It uses the gravimetric method specified in ISO 8655-6[2] as the reference method for calibrating piston-operated volumetric apparatus. It has been arranged in paragraphs to facilitate direct access to different aspects of this kind of evaluation as follows:
— modelling the measurement by describing the physical equations necessary to calculate the volume using the gravimetric method of measurement;
— determination of the standard uncertainty of measurement associated with the volume V 20 by describing the calculation procedure according to the GUM;
— determination of the sensitivity coefficients with an example of the calculation of all sensitivity coefficients by using complete equations, approximations of equations and by giving numerical values for standard conditions;
— determination of the standard uncertainty associated with the volume delivered by a piston-operated volumetric apparatus giving the combination of the standard uncertainty associated with the volume V 20 measured using the gravimetric measuring system and the experimental standard deviation associated with the volume delivered by the a...